Missing: specimen needs to be assgined to the observation

    XMLWordPrintableJSON

Details

    • Icon: Change Request Change Request
    • Resolution: Unresolved
    • Icon: Medium Medium
    • EU Laboratory Report (FHIR)
    • 0.1.0-ballot [deprecated]
    • HL7 Europe
    • Observation Results: laboratory

    Description

      At least one specimen has to be assigned to an observation.

      The current assignment of 0..1 specimen to a diagnostic report is not correct, the reference should not happen in the diagnostic report but in the observation.

      Explanation: within the diagnostic report could exist more than one specimen. For every single observation must be clear, from which specimen the measurement is taken.

      Kind regards from Martina Sender (mio42 GmbH, Germany)