Uploaded image for project: 'FHIR Specification Feedback'
  1. FHIR Specification Feedback
  2. FHIR-43708

Add to DeviceMetric - alertDetection

    XMLWordPrintableJSON

Details

    • Icon: Change Request Change Request
    • Resolution: Persuasive
    • Icon: Medium Medium
    • FHIR Core (FHIR)
    • R6
    • Devices
    • DeviceMetric
    • Hide

      Motion to accept issue as PERSUASIVE and to update DeviceMetric accordingly.

      (Cooper / Silver, 10/0/1, 2024.01.30)  

       

      Note:  DeviceMetric updates should not be applied until after approval of the DeviceAlert new FHIR resource proposal.

      Show
      Motion to accept issue as PERSUASIVE and to update DeviceMetric accordingly. (Cooper / Silver, 10/0/1, 2024.01.30)     Note:  DeviceMetric updates should not be applied until after approval of the DeviceAlert new FHIR resource proposal.
    • Enhancement
    • Compatible, substantive
    • R6

    Description

      DEV WG / PSS-2005 Device Alerting Project has decided to update the DeviceMetric resource (R6) with a new data element:  alertDetection.  This is consistent with a similar update to the Device resource (see FHIR-43660) and the new [DeviceAlert FHIR resource being proposed|https://confluence.hl7.org/display/FHIR/DeviceAlert+FHIR+Resource+Proposal]

      Proposed data element structure (see structure mockup page)

       ** 

      Name Flags Card. Type Description and Constraints Long Definition / Implementer Comments
      alertDetection   0..* BackboneElement Describes the alert detection activation states and other relevant information.  If code is not present, activationState represents the overall state for the device metric. Additional entries may provide deviating state for specific alerts.
      If code is not present, limitRange applies to DeviceMetric.type. For metrics with multiple components (e.g., blood pressure measurement) the code identifies the component.
      • code
        1..1 CodeableConcept Alert code for specific alert activation state 
      all | high respiration rate | sensor disconnected 
      binding:FHIR Alert Code Value Set ( example )
       
      • activationState
        1..1 code on | off | paused
      binding:FHIR Alert Activation State Value Set ( required )
       
      • priority
        0..1   high | medium | low | info May omit if it applies to all priorities
      • limitRange
        0..1 Range Alarm limits that are applied in the determination of the alert condition  Alert limits may be upper, lower or both.   Note:  If units are present in the Range, they should be the same as in the DeviceMetric.  If units are not present in the Range, they are are assumed to be the same as in the DeviceMetric.

       

       

      Attachments

        Activity

          People

            Unassigned Unassigned
            todd.cooper Todd Cooper
            Watchers:
            3 Start watching this issue

            Dates

              Created:
              Updated:
              Resolved: