Add Device.Operation.mode value set

    XMLWordPrintableJSON

Details

    • Icon: Change Request Change Request
    • Resolution: Persuasive with Modification
    • Icon: Medium Medium
    • FHIR Core (FHIR)
    • R5
    • Orders & Observations
    • Device
    • Hide

      Add Device.operationState.mode value set as follows:

      Code Display Definition
      normal Normal The device operates in a mode that supports the fulfillment of its clinical functions.
      demo Demo The device operates in a mode that is intended for demonstration purposes only. Arbitrary values are generated.
      service Service The device operates in a mode that is intended for correcting an functional problem of the device only. Arbitrary values may be generated.
      maintenance Maintenance The device operates in a mode that is intended for preventative and/or scheduled maintenance purposes only.
      test Test The device operates in a test mode that is not intended to be used for production/operational purposes.
      Show
      Add Device.operationState.mode value set as follows: Code Display Definition normal Normal The device operates in a mode that supports the fulfillment of its clinical functions. demo Demo The device operates in a mode that is intended for demonstration purposes only. Arbitrary values are generated. service Service The device operates in a mode that is intended for correcting an functional problem of the device only. Arbitrary values may be generated. maintenance Maintenance The device operates in a mode that is intended for preventative and/or scheduled maintenance purposes only. test Test The device operates in a test mode that is not intended to be used for production/operational purposes.
    • Marti Velezis / John Rhoads: 8-0-0
    • Enhancement
    • Compatible, substantive
    • R5

    Description

      Add Device.Operation.mode value set as follows:

      Code Display Definition
      normal Normal The device operates in a mode that supports the fulfillment of its clinical functions.
      demo Demo The device operates in a mode that is intended for demonstration purposes only. Arbitrary values are generated.
      service Service The device operates in a mode that is intended for services purposes only. Arbitrary values may be generated.
      maintenance Maintenance The device operates in a mode that is intended for maintenance purposes only.