    • Hide

      Note: This comment was previously disposed of as "Deferred" or "Consider for Future Use" it has been reopened for reconsideration for this ballot cycle

      Will add search combination of patient + status as a recommended search requirement. (SHOULD).

      Show
      Note: This comment was previously disposed of as "Deferred" or "Consider for Future Use" it has been reopened for reconsideration for this ballot cycle Will add search combination of patient + status as a recommended search requirement. (SHOULD).
    • Brett Marquard/Juliet Rubini: 17-0-2
    • Enhancement
    • Non-substantive

    Description

      The comment was submitted on behalf of FDA CDRH:


      The search string provided (GET [base]/Device?patient=[id]) is imprecise/does not meet the MU2015 needs to provide the "UDI of implantable devices"..

      GET [base]/Device?patient=1032702 will return all the devices associated with Patient resource with id=1032702 *. In order ot meet the MU2015 requiremetns, client applications/requesters need to be able ot search for "active" implants to obtain an *implantable device list. This type of query will return any devices current/active and past/inactive including durable medical equipment like wheel chairs, walkers as well as prostheses, corrective lenses, etc.

      Attachments